Vitrek 95 Series - PULSE Configuration Fields

Vitrek 95 Series
154 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
95x Series Operating Manual - June 27, 2016
Page 143 of 154
3
<NR3>
Delay time (in seconds)
4
<NR3>
Minimum Leakage Current Limit (in A)
5
<NR3>
Maximum Leakage Current Limit (in A)
6
<NR1>
Time period for arc detection (4, 10, 15, 20, 30 or 40)
<EMPTY>
Arc detection disabled
7
<NR1>
Arc detection limit (in mA)
<EMPTY>
Only valid if arc detection disabled in field 6
8
<STRING>
ABORT - Abort on failure
CONT - Continue on failure
Example –
ADD,DCI,5.0,0.0,0.0,50e-6,,,ABORT
Configures the following (in order) –
A DCI type test step
5 second test time
0 second delay time
No minimum leakage current
50uA maximum leakage current
No arc detection (so following field is also <EMPTY>)
Abort test sequence if fails
PULSE CON FIGURATION FIELDS
Field #
Field Format
Value
1
<STRING>
PULSE
2
<NR3>
Level (in V)
3
<STRING>
POSITIVE - Unipolar +ve
NEGATIVE – Unipolar –ve
BIPOLAR - Bipolar
4
<NR3>
Ramp time (tr) (in seconds)
5
<NR3>
Hold Time (th) (in seconds)
6
<NR3>
Breakdown Current Limit (in A)
7
<NR1>
Time period for arc detection (4, 10, 15, 20, 30 or 40)
<EMPTY>
Arc detection disabled
8
<NR1>
Arc detection limit (in mA)
<EMPTY>
Only valid if arc detection disabled in field 7
9
<STRING>
ABORT - Abort on failure
CONT - Continue on failure
Example –
ADD,PULSE,250.0,POSITIVE,1e-3,10e-3,0.05,4,10,ABORT
Configures the following (in order) –
A PULSE type test step
A single positive pulse
1ms ramp time