Part number 20055, Rev 0
1
SOLID DOOR LABORATORY PLUS SERIES FREEZERS
SCPMF-2020, SCPMF-2030, SCPMF-3020, SCPMF-3030, SCPAF-3020
OWNER’S INSTRUCTIONS
This manual describes how to operate and care for your appliance to get the best, most efficient, performance.
READ THIS BOOK!
Note to Customer:
This merchandise was carefully packed and thoroughly inspected before leaving our plant. Responsibility
for its safe delivery was assumed by the carrier upon acceptance of the shipment. As directed on the
side of your packing carton, claims for loss or damage sustained in transit must be made on the carrier as
follows:
A.) Visible Loss, Damage, Shortage External Evidence of Loss or Damage: This type of damage
must be noted on the freight bill and acknowledged by the carrier’s agent (driver) at time of delivery.
Make sure you get a signed copy. Send a written request for an inspection to the carrier.
B.) Concealed Damage: This type of damage may not be discovered until the unit is being unpacked.
When concealed damage is discovered, stop unpacking immediately and contact the carrier
immediately to report the claim and request an inspection. This should be done as soon as possible
and, in any case, must be done within 15 days or receiving the merchandise. If at all possible, do not
move the item and save all packaging material for carrier’s inspection.
C.) FAILURE TO FOLLOW THESE INSTRUCTIONS MAY RESULT IN THE CARRIER REFUSING TO
HONOR YOUR COMPANY’S CLAIM. UNDER NO CIRCUMSTANCES SHOULD THE
MERCHANDISE BE RETURNED TO THE MANUFACTURER. NO RETURNS WILL BE ACCEPTED
WITHOUT PRIOR AUTHORIZATION.
Horizon Scientific
125 Varnfield Drive
Summerville, SC 29483
Phone: 800-648-4041
Fax: 843-821-8051
Warranty:
Two-Year Parts and Labor
5 Years Compressor Parts
Revision Date
08/23/2012

Questions and Answers

  • D
    David JacksonAug 8, 2025
    What to do if my VWR SCPMF-2020 Freezer does not run?
    • E
      Elizabeth GarciaAug 8, 2025
      If your VWR Freezer isn't running, start by checking the electrical circuit to ensure it's 110-120V 60Hz. Next, make sure the power cord is securely plugged in. Verify that the electrical outlet has power by checking for tripped breakers or blown fuses. Finally, ensure the unit is not plugged into a Ground Fault Circuit Interrupter (GFCI) outlet.
  • T
    Terry LaneAug 12, 2025
    Why is my VWR SCPMF-2020 not maintaining the proper temperature?
    • E
      Erik FernandezAug 12, 2025
      If your VWR Freezer isn't maintaining the correct temperature, first ensure the room temperature is between 65°F and 85°F. Check that the door is closing properly and that the stored product isn't overloaded, distributing items evenly on each shelf. Also, make sure products aren't pushed against the rear wall. If the evaporator is blocked by frost or ice, remove the products, unplug the freezer, and allow it to defrost. Finally, ensure any third-party thermometer is placed correctly in the middle of the freezer.
  • J
    Joel AllenAug 16, 2025
    Why does my VWR Freezer run too long?
    • V
      Veronica JohnstonAug 16, 2025
      If your VWR Freezer is running for too long, try these steps. Minimize how often you open the door. Adjust the temperature control to a warmer setting. Finally, try to reduce the room temperature.
  • M
    Mark HardinAug 20, 2025
    Why does moisture collect inside my VWR SCPMF-2020?
    • T
      Tina RobersonAug 20, 2025
      If moisture is collecting inside your VWR Freezer, inspect the door gasket for debris, cracks, or obstructions. Ensure the freezer isn't facing a doorway or under an air conditioning vent; relocate it or redirect the vent if necessary. Minimize door openings and be aware that hot, humid weather increases condensation. Also, confirm there's a water trap (U-shaped loop) in the drain tube near the compressor.
  • C
    Charles TorresAug 25, 2025
    How to fix a VWR Freezer door that will not close?
    • C
      Cynthia BrownAug 26, 2025
      If the door of your VWR Freezer won't close, see the Leveling section at the beginning of this manual. Also check the door seal for dirt, debris, or obstructions.
  • D
    Dean MillerAug 29, 2025
    Why does moisture collect on the outside of my VWR Freezer?
    • J
      Jasmine ThomasAug 29, 2025
      If moisture collects on the outside surface of your VWR Freezer, be aware that hot, humid weather increases condensation. The moisture should disappear as humidity decreases.
  • A
    Angela KlineSep 2, 2025
    How to get rid of odor inside VWR SCPMF-2020 Freezer?
    • M
      Michael ChangSep 2, 2025
      If there is an odor inside your VWR Freezer, see the section on maintenance and cleaning in this manual. Make sure product containers are tightly sealed to prevent leakage.

VWR SCPMF-2020 Specifications

General IconGeneral
BrandVWR
ModelSCPMF-2020
CategoryFreezer
LanguageEnglish