Error Messages
Display
Messages
Corrective Action
E1
Appears when either the up or down
If the E1 message appears even when
no keys are being pressed, replace
control.
E2
Check all settings and correct if
necessary.
EP
Appears when the probe is open,
Check to see if the sensed
temperature is out of range. If not,
check for probe damage by comparing
it to a known am
bient temperature
between
-
30°F and 220°F. Replace the
probe if necessary.
EE
Appears if the EPROM data has
This condition cannot be field repaired.
Replace the control.
CL
Remove power to the control for at
least five seconds. Reapply power. If
the CL message still appeared, replace
the control.