3
SAFETY INFORMATION
Cleaning
Wipe the case with a damp cloth and mild detergent. Do not use abrosives
or solvents. Dirt or moisture in the terminals can affect readings.
Safety: Conforms to IEC 61010-1 (EN 61010-1), IEC 61010-2-032
(EN 61010-2-032), CAT III 1000V, CAT IV 600V, Class II, Pollution
degree 2 Indoor use.
CAT III: Is for measurements performed in the building installation.
CAT IV: Is for measurements performed at the sourec of the low-voltage
installation.
EMC: Conforms to EN 61326-1: 2006
The symbols used on this instrument are:
Dangerous voltage.
Caution, refer to accompanying documents
Equipment protected throughout by Double
insulation (Class II)
Alternating current
Direct current
Ground
2609C.p65 ?2009/12/15, 上午 09:155