CLX-XT (28652) Page 21
REV 5 (5/2020)
Alarm 1 Set Point: This prompt is used to select the set point for this alarm; this
is indicated by “S/P” shown on the lower row of the display. Select the desired
alarm level by using the and buttons. Once the desired set point has been
set, press the button to accept it.
7.4.2 Alarm 2
Repeat the procedure listed in section 7.4.1 to set up the parameters for alarm 2. If
a selection was made to turn the alarm OFF, the next selection for the speed of
response RESP is shown.
The use of the Remote Standby Option 25264 limits the relays to a single Normally
open connection. Follow Alarm 1 Instructions.
Due to the cyclic nature of the CLX-XT, relay chatter is not an issue. There is no
need for alarm delays or hysteresis.
7.5 Enabling the Security Access
The instrument is equipped with a security access. If this option is turned on, the user is
required to input the access code into the instrument to
get to any mode other than AUTO. The only code is
333. This code may not be changed. See section 5.2 for
more information on this security feature. The security
key icon will be visible and flashing on the display
whenever the access option is selected using the or
buttons. (On or OFF).
7.6 Extended Settings
The last settings are grouped together to prevent them
from being adjusted by accident. To gain access to the
extended settings, select On using the or buttons
and press the button.
If extended settings are set to OFF, pressing the button will save all settings and the
CLX-XT will automatically return to the normal AUTO mode of the instrument.