CLX-XT (28652) Page 24
REV 5 (5/2020)
7.13 4mA Adjustment
If the 4-20 mA setting is turned ON (7.1 Setting the 4-20 mA Output), the following two
menus will appear. The first menu outputs a constant 4
mA while allowing for a small amount of adjustment.
The adjustment can be made using the and buttons.
This adjustment will allow the operator to make the
CLX-XT agree with a PLC or SCADA system. The
adjustment limits are ± 200 counts or about ± 0.2 mA.
This setting will be slightly different on each instrument as each CLX-XT will be factory
set to 4.00mA. Press the button when adjustments are complete to save this setting and
move on to the 20mA adjustment.
7.14 20mA Adjustment
This menu operates similar to the previous menu. This menu outputs a constant 20 mA
while allowing for a small amount of adjustment. The adjustment can be made using the
and buttons. The adjustment limits are ± 1000 counts or about ± 1 mA.
This setting will be slightly different on each instrument
as each CLX-XT will be factory set to 20.00mA.
When complete with the 20mA adjustment, press the button to exit to AUTO mode and
save all configuration settings.