CLX-XT (28652) Page 29
REV 5 (5/2020)
9.4 Reagent Clogs
If reagents fail to flow or Prime it may be due to a clog in either the tubing or at a check
valve. To alleviate this you may have to flush the system with Chlorine Free water,
preferably Deionized water. See section10.3 Check Valve Flushing Kit.
9.5 Diagnostic Chart
Check wiring. See sections
4.3.4 and 7.2
(1) Install bubble elimination
chamber.
Readings are lower than expected
measurement cuvette
(2) Measurement cuvette
dirty
(3) Reagents bad or expired
(4)
dispensed
(1) Install desiccant cartridge
kit
(2) Replace or clean cuvette
(3) Replace reagents
(4) Check buffer lines and
Check sample. Sample may be
too high to read.
Upper display shows nOnE while
attempting to calibrate
No current reading displayed
Wait for CLX-XT to post a
reading