Logistical requirements
Oxygen supply: at least a 2-l bottle, lled ....................................................... Check
Emergency medical team familiar with NIV ..................................................... Check
Clinical requirements
alert, cooperative (GCS >12) ......................................................................... Check
breathing spontaneously ................................................................................ Check
if applicable, light sedation of agitated patients e.g. morphine
(5-10 mg i.v. titrated) or a short-acting benzodiazepine.
Indications
Dyspnea ......................................................................................................... Check
Respiratory rate >25/min (count!) ................................................................. Check
SpO
2
<90% despite O
2
administration ........................................................... Check
Contraindications
Absolute contraindications: ............................................................................ Check
absence of spontaneous respiration, gasping, airway obstruction,
gastrointestinal bleeding or ileus
Relative contraindications: .............................................................................. Check
Coma, massive agitation, hemodynamic instability, severe hypoxemia
(SpO
2
<75% despite O
2
), problems with airway access, status post
gastrointestinal surgery
Pulmonary edema
Primary device settings
Ventilation mode: ............................................................................................CPAP
PEEP (according to comfort and oxygenation): ......................................5/7/10mbar
FiO
2
: ..............................................................................................Air Mix or 100%
Aim and success criteria
Target SpO
2
: >90% ....................................................................................... Check
Decrease in dyspnea ...................................................................................... Check
Falling respiratory and heart rate .................................................................... Check
If applicable, improved vigilance ..................................................................... Check
NIV sequence: Set the device, place the mask on the patient’s face (explain the
measure!), connect the mask to the hose system while the device is running. Aim:
Synchronization of patient and device
SOP
(Standard Operating Procedure)
Non-invasive ventilation (NIV) by the emergency medical
services

Other manuals for Weinmann MEDUMAT Standard2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUMAT Standard2 and is the answer not in the manual?

Weinmann MEDUMAT Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUMAT Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals