Changing the ventilation mode
Requirements:
The device is switched on.
A ventilation mode is set.
1. Briefly press the menu button .
2. Select the Mode field.
3. Select a ventilation mode.
4. Adjust the ventilation parameters and press start.
Ventilating the patient
Selecting emergency mode
1. Briefly press the On/Off button .
2. Select emergency mode:
Emergency Pediatric
Emergency Child
Emergency Adult
Selecting previous patient
1. Briefly press the On/Off button .
2. Select Previous patient field.
RSI mode
Performing RSI ventilation
Requirements:
The device is switched on.
The MEDUtrigger is connected to the device.
RSI was set as the ventilation mode for a
new patient. The Demand function is
automatically activated.
1. For the function Manual, navigate to the
field Demand.
2. Select the Manual field.
3. To perform continuous ventilation
following successful airway management,
select the Contin. field. The device
switches to IPPV mode.
Selecting a new patient
1. Briefly press the On/Off button .
2. Select New patient field.
3. Select height and gender or navigate to the
Height field and turn the navigation knob further
to select the desired patient group.
4. Press the navigation knob next.
5. Select a ventilation mode.
6. If necessary: Adjust the ventilation parameters and
press start.
Pausing the ventilation
Requirements:
Cardiac rhythm analysis or defibrillation
is being performed.
The device is switched on.
The CPR mode is set.
Continuous ventilation
is activated.
1. Select the Pause field.
The ventilation is paused for 50 seconds.
CPR mode
Performing CPR ventilation
Requirements:
The device is switched on.
The MEDUtrigger is connected to the device.
A ventilation mode for a new patient is set.
1. Briefly press the CPR button
.
The device switches to the mode CPR 30:2/CPR 15:2/CPR IPPV (depending on
the preset). The green LEDs on the MEDUtrigger light up.
2. If necessary:
Change algorithm:
30:2
15:2
Continuous
3. Press and hold the MEDUtrigger button
during the ventilation phase until two
mechanical breaths are performed.
4. To exit the CPR mode,
press the CPR button
again.
To enable the selection of the function Manual, a MEDUtrigger must
be connected. Otherwise, this function will not be displayed.
If you do not select a menu within 20 seconds, the device automati-
cally switches to the Previous patient menu and immediately begins
ventilation of the patient. In this case, the device uses the ventilation mode
and ventilation parameters set for the last patient to be ventilated. To switch
off the countdown before the 20 seconds have elapsed, move one of the
navigation knobs.
MEDUMAT Standard
2
MEDUMAT Standard
2
Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUMAT Standard 2 and is the answer not in the manual?

Weinmann MEDUMAT Standard 2 Specifications

General IconGeneral
Device TypeVentilator
Tidal Volume Range50 - 2000 ml
PEEP0 - 20 mbar
TypeTransport Ventilator
ApplicationEmergency and Transport
Ventilation ModesIPPV, SIMV, CPAP
Inspiratory Time0.5 - 3 s
Power Supply100 - 240 V AC

Summary

Patient Ventilation Setup and Modes

Patient Selection

Select emergency, previous, or new patient and initial parameters.

Ventilation Modes

Perform CPR or RSI ventilation using the MEDUtrigger.

Ventilation Control

Change ventilation mode, adjust parameters, or pause ventilation.

Device Operation and Power Management

Device Power

Switch the device on and off using the On/Off button.

Function Check

Perform a pre-use function check for device readiness.

Battery Charging

Charge the battery in the device or in a charging station.

Status Indicators

Interpret line power and battery status indicators.