98 Replacement parts Welch Allyn VSM 300 Series
Notes
Order
number
Description Notes
Special service tools
06138-000 Temperature test key (tests monitor only)
01802-110 9600 Plus Calibration Tester, 110 V (tests temperature probe
and monitor)
01802-220 9600 Plus Calibration Tester, 220V (tests temperature probe
and monitor)
020-0702-XX Fixture, cuff simulator assembly
407590 Battery substitution connector
810-1784-XX Welch Allyn monitor service utility application CDROM 4
020-0660-XX SW update CDROM for POEM1 update to version 3.20.00
008-0842-00 Serial cable, utility
VSM 300 Series AC power adapter
AC power adapter (8V, 0.75A, 120 VAC USA) 7
AC power adapter (8V, 0.75A, 240 VAC EU) 8
AC power adapter (8V, 0.75A, 240 VAC NZ, Australia) 8
1 Requires Welch Allyn monitor service utililty application (810-1784-XX) and
serial cable (008-0842-XX) when the replacement board software version is
different from the existing board.
3 Use with main board 031-0150-03 and higher
4 Used to characterize NIBP, read log files, and update NIBP, SpO2, and temp
modules. Requires serial cable (008-0842-XX).
5 Requires Welch Allyn Service Center installation and programming.
6 Use 020-0634-XX to replace PCB assembly 031-0150-01. Use the latest
version of 031-0150-XX to replace 031-0150-03 or higher.
7 Refer to the North America price list for part numbers.
8 Refer to the International price list for part numbers.
9 Field replacement of the temperature board requires main board
031-0150-03 or higher. Devices with main board 031-0150-01 must be
returned to a Welch Allyn Service Center for temperature board
replacement.