12-16-08
Exx
When an error code consisting of an “E” and a two digit number xx = number)
appears on the audiometer’s display, a system error has been detected. The AM 232
Manual Audiometer will enter a “lockout” mode which will not permit the instrument
to operate. The specific error code will remain on the display for several seconds,
then the instrument will shut itself down completely. Should an Exx appear on the
LCD, take the following steps:
1. Power down, power up again. This could be only a temporary failure and may
never appear again. However, should the Exx message appear again, proceed to
steps b and c.
a. Write down the numbers displayed on the display.
b. Contact your certified service representative and give them the numbers you
recorded.