to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The device is suitable for use
in all establishments,
including domestic
establishments and those
directly connected to the
public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
Harmonic
emissions
IEC61000-3-2
Class A
Voltage
Fluctuations /
Flicker
Emissions IEC
61000-3-3
Complies
Guidance and Declaration - Electromagnetic
Immunity
The Health Monitor is intended for use in the
electromagnetic environment specified below. The
customer or the user of the Health Monitor should
assure that it is used in such an environment.
Immunit
y test
IEC60601
test level
Compli
ance
level
Electromagnetic
environment -
guidance
Electrost
atic
discharg
e (ESD)
IEC
61000-4-
2
± 6 kV
contact
±
8 kV air
± 6 kV
contact
±
8 kV
air
Floors should be
wood, concrete
or ceramic tile. If
floors are
covered with
synthetic
material,
the relative
humidity should
be at
least 30 %.
Electrical
± 2 kV for
± 2 kV
Mains power