25
power of
transmitte
r (W)
80 MHz
Pd 2.1
800
MHz
Pd 2.1
2.5 GHz
Pd 3.2
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.20
1.20
2.30
10
3.80
3.80
7.30
100
12.00
12.00
23.00
For transmitters rated at a maximum output power
not listed above, the recommended separation
distance d in metres (m) can be estimated using the
equation applicable to the frequency of the
transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation
distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all
situations. Electromagnetic propagation is affected
by absorption and reflection from structures,
objects and people.