GENERAL INFORMATION
R600A Sealed System 2017
n
1-5
Gas detector Continued
5. The frequency of the ck (sound) is an indicaon of the
sensivity. Move the knob unl the cking is rapid, for High
sensivity, or slow for Low sensivity.
NOTE: If a steady ck cannot be maintained, it is indicave that
the baeries may need to be recharged.
6. Search the general area of the suspected refrigerant leak.
When a detectable compound enters the p, the ck rate
speeds up.
7. In conjuncon with the increased speed of the ck the LEDs
will light from le to right as a combusble is detected. The
larger the concentraon, the more the LEDS will light.
8. In most cases, it will not be necessary to adjust the sensivity
of the detector. However, if the siren sounds before a possible
leak source can be found, it is likely that air is contaminated with
heavy concentraons of gas. Therefore, you may desensize the
detector by moving the adjustment knob down to low sensivity
unl the cking slows to normal.
9. If your are searching for extremely small leaks, Make certain
the control knob is in the High posion or you hear rapid cking.
Ulize vacuum pumps, recovery equipment, and other tools
that are rated for and cered for use with A3 ammable
refrigerants. Vacuum pumps must be cered for the
refrigerant being used.
Evacuation:
If a vacuum pump rated for A3-refrigerant is not available,
check with your vacuum pump manufacturer for
instrucons about using equipment for A3s such as R600a.