Questions and Answers

  • C
    Calvin IngramAug 11, 2025
    What to do if the Whitehall Para-Therapy PT-6-S paraffin wax is too hot?
    • T
      Tammy GreenAug 11, 2025
      If the paraffin in your Whitehall Medical Equipment is too hot, it could be due to a couple of reasons. First, the paraffin level might be too low; in this case, add more paraffin to the unit. Alternatively, the thermostat might be broken and need replacement.
  • J
    Jonathan DanielsAug 17, 2025
    Why is the paraffin wax not melting and the indicator light is off on my Whitehall Para-Therapy PT-6-S?
    • C
      christopher38Aug 17, 2025
      If the paraffin wax is not melting and the indicator light is off on your Whitehall Medical Equipment, there could be a few potential causes: * The unit might not be receiving power. Check that the unit is properly plugged into the wall outlet and securely attached to the appliance inlet. * There might be no power from the wall outlet. Inspect the fuse box for the wall outlet. * The mains fuse might be burnt. If so, replace the burnt fuse.
  • Z
    zwalkerAug 23, 2025
    What to do if the paraffin wax is not melting but the indicator light is on in my Whitehall Medical Equipment?
    • S
      Sara GarzaAug 23, 2025
      If the paraffin wax is not melting, but the indicator light is on in your Whitehall Medical Equipment, it could be due to: * A burnt or faulty thermal cutout. Check the thermal cutout for continuity with a multi-meter; if there's no continuity, replace the thermal cutout. * A broken thermostat. In this case, replace the thermostat. * A burnt or faulty heating element. Replace the heating elements.
  • A
    Andre BowersAug 29, 2025
    What to do if the paraffin is solidifying at the top surface in my Whitehall Medical Equipment?
    • D
      David BarberAug 30, 2025
      If the paraffin is solidifying at the top surface in your Whitehall Medical Equipment, it's likely that the paraffin is low on oil. Add paraffin oil (food grade mineral oil) to the paraffin.
  • A
    Abigail MooreSep 4, 2025
    How to fix paraffin is melting but the indicator light is off on Whitehall Medical Equipment?
    • J
      Jacqueline WrightSep 5, 2025
      If the paraffin is melting but the indicator light is off on your Whitehall Medical Equipment, the indicator light is likely burnt out. Replace the indicator light.