3. Safety
3.1 Intended use
This CPG1200 digital pressure gauge can be used as a calibration instrument and
for any application (indoor/outdoor) that requires highly accurate measurement of
pneumatic or hydraulic pressures.
Measuring ranges between -1 ... 1,000 bar [-14.5 ... 15,000 psi] can be measured.
The CPG1200 must only be operated with media of fluid group 2 in accordance with
directive 2014/68/EU article 13, which are considered harmless for the wetted parts
across the entire area of application of the instrument. Do not use the CPG1200 with
abrasive and viscous media or oxygen.
Use with oxygen is possible as an option. In this case, please contact WIKA.
For contact details, see chapter 1 “General information” or the back page of
the operating instructions.
The use of unstable fluids, especially hydrogen, must be avoided.
If the CPG1200 is used in applications with oil as a pressure medium, make sure it
will not be used with combustibles or gases directly after that because it can lead to
dangerous explosions and danger to personnel and machinery.