GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC EMISSIONS
The Willow Go™ Breast Pump is intended for use in the electromagnetic environment specied below. The customer or user of the Willow Go™ Breast Pump
should ensure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment –Guidance
Radiated RF EM Fields
IEC 61000-4-3
10V/m 80MHz to
2.7 GHz
10V/m 80MHz to
2.7 GHz
Portable and mobile RF Communications equipment should be used no closer to
any part of the Willow Go™ Breast Pump, including cables, than the recommended
separation distance calculated from the equation applicable to the frequency of the
transmitter.
Recommended separation distance:
d = 1.2√P 80 MHz to 800 MHz
d = 1.2√P 800 MHz to 2.7 GHz
Where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from xed RF transmitters, as determined by an electromagnetic site
survey,
a. should be less than the compliance level in each frequency range.
b. Interference may occur in the vicinity of equipment marked with the following
symbol:
Conducted
disturbances induced
by RF elds
IEC 61000-4-6
3 V; 150 kHz to 80
MHz
6 V; ISM and amateur
radio bands between
0.15 MHz and 80 MHz
Not applicable The Willow Go™ Breast Pump is exempt from this test.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reection from structures, objects, and people.