6
C-QIM-2731-T01_EN_Rev.2
Tel +82-2-300-2300
Fax +82-2-300-2400
webmaster@wisecontrol.com
1) Thermocouple (TC): Class 1,2,3(IEC/EN 60584-2) or Standard / Special(ASTM E230, ANSI MC 96.1)
Type
Class
Temperature range
Maximum deviation
K
Standard
-200 ~ 0
The greater of ±2.2 or ±2%
0 ~ 1260
The greater of ±2.2 or ±0.75%
Special
0 ~ 1260
The greater of ±1.1 or ±0.4%
E
Standard
-200 ~ 0
The greater of ±1.7 or ±1%
0 ~ 870
The greater of ±1.7 or ±0.5%
Special
0 ~ 870
The greater of ±1.0 or ±0.4%
B
Standard
870 ~ 1700
±0.5%
Special
870 ~ 1700
±0.25%
R
Standard
0 ~ 1480
The greater of ±1.5 or ±0.25%
Special
0 ~ 1480
The greater of ±0.6 or ±0.1%
S
Standard
0 ~ 1480
The greater of ±1.5 or ±0.25%
Special
0 ~ 1480
The greater of ±0.6 or ±0.1%
N
Standard
0 ~ 1260
The greater of ±2.2 or ±0.75%
Special
0 ~ 1260
The greater of ±1.1 or ±0.4%
T
Standard
-200 ~ 0
The greater of ±1.0 or ±1.5%
0 ~ 370
The greater of ±1.0 or ±0.75%
Special
0 ~ 370
The greater of ±0.5 or ±0.4%
J
Standard
0 ~ 760
The greater of ±2.2 or ±0.75%
Special
0 ~ 760
The greater of ±1.1 or ±0.4%
*Note.
Each model has different certification types. See “3.3 Temperature sensor types” and look up the
certifications for the applicable type in the table.
2) Resistance thermometer detector (RTD): Class A, B (IEC/EN 60751)
Tolerance
class
Temperature range of validity (°C)
Tolerance value ()
For film resistors
A
-30 to +300
± ( 0.15 + 0.002 | t | )
B
-50 to +500
± ( 0.3 + 0.005 | t | )

Table of Contents