Conducted RF disturbances
Conducted RF
IEC 61000-4-6
3 V 150 kHz - 80 MHz
6 V in ISM and radio
amateur band
3 V
6 V
Portable and mobile radios should not be used at a
distance of less than 30 cm from the device,
including the lines.
Radiated RF disturbances
Radiated RF
IEC 61000-4-3
Proximity fields
10 V/m 80 MHz 2,7 GHz
27 V/m 385 MHz
28 V/m 450 MHz, 810 MHz,
870 MHz, 930 MHz,
1720 MHz, 1845
MHz, 1970 MHz,
2450 MHz
9 V/m 710 MHz, 745 MHz,
780 MHz, 5240
MHz, 5500 MHz,
5785 MHz
10 V/m
27 V/m
28 V/m
9 V/m
The field strength of stationary radio transmitters
should be less than the compliance level at all
frequencies according to an on-site investigation.
Interference is possible in the vicinity of devices
bearing the following symbol
The acceptance criterion for EMC testing is basic safety for the operator, third parties and the environment as described in the basic standard IEC
60601-1.
12. Disposal
The packaging of the device is made of recyclable materials and can be reused or disposed of with
waste paper.
Protect the environment!
The HS-100 and the power supply unit contain valuable materials and must be
disposed of separately from electrical and electronic equipment in accordance with
Directive 2012/19/EU.
The device or parts of the device must not be disposed of in household waste!
Please contact your local authorities and collection points regarding the disposal of
potentially biologically contaminated components.