Seite 8 von 100
2.2 INITIAL STARTUP OF THE INSTRUMENTS
General remarks:
The instruments belonging to a measuring group as described below in Pt. 2.4 is already done at WYLER’s
when the instruments are delivered.
When a group of instruments are shut off e.g. after a measurement is completed, the group set-up remains
saved. When started again, the communication is immediately ready, no additional set-up is required.
When a group of instruments is extended e.g. by joining or replacing an instrument, this newly added
instrument must be joined according Pt.2.4 „Combine a group....“
First time start-up of a measuring group (BlueLEVEL and BlueMETER) with radio transmission (wireless
transmission of the data):
1. Remove the instruments with care from the transport and storage case and insert the batteries
2. Place the instruments on a clean horizontally aligned surface and switch on the instruments
(BlueLEVEL and BlueMETER). Keep the two keys <ON/MODE> and <ZERO/SELECT ±> (BlueLEVEL),
respectively the key <ON/MODE> (BlueMETER) pressed until all 5 LED’s are on. When the keys are
released on the display is seen „SYSTEM TEST“.
If the keys <ON/MODE> and <ZERO/SELECT ±> (BlueLEVEL), respectively the key <ON/MODE>
(BlueMETER) are pressed longer than 10 seconds when STARTING the instrument all the LED’s
are beginning to blink and the automatic shut-off system is deactivated. In the standard mode the
instrument is automatically shut off after 60 minutes.
3. The instruments are now communicating and after a few seconds the measured values are displayed.
The battery power should be checked (see below)
4. The measuring task may now be started.
First time start-up of a measuring group (BlueLEVEL and BlueMETER) without radio transmission (transmission of
the data by cables):
1. Remove the instruments with care from the transport and storage case and insert the batteries
2. Place the instruments on a clean horizontally aligned surface and connect the instruments with the
cables supplied
3. Switch on the instruments. Keep the two keys <ON/MODE> and <ZERO/SELECT ±> (BlueLEVEL),
respectively <ON/MODE> (BlueMETER) pressed until all 5 LED’s are on. When the keys are released
on the display is seen „SYSTEM TEST“.
If the keys <ON/MODE> and <ZERO/SELECT ±> (BlueLEVEL), respectively <ON/MODE>
(BlueMETER) are pressed longer than 10 seconds when STARTING the instrument all the LED’s
are beginning to blink and the automatic shut-off system is deactivated. In the standard mode the
instrument is automatically shut off after 60 minutes.
4. The instruments are ready for use and the measured values are displayed. The battery power should be
checked (see below)
5. The measuring task may now be started.
Remarks concerning the battery power:
Display of the
actual battery
power
(Example: 2,6 V)
Display of the actual
battery power
(Example: 3,6 V)
The lowest voltage is 1,7 Volt. After this limit has
been reached a battery symbol blinking is
displayed. The batteries should now be replaced in
due time.
The lowest voltage is 2.8 Volt. After this limit has been
reached a battery symbol blinking is displayed. The
batteries should now be replaced in due time.