3.5. Function SEND (Print-Function)
Using the key
SEND/ESC will send the displayed value through the port "OUT" to a connected PC, or Laptop
via the RS232 port. This function can also be used in combination with HOLD in order to transmit the „frozen“
value.
The function SEND may be activated from a connected PC/Laptop by transmitting a letter „P“ (as letter) through
the serial port RS 232
Data format OUT-Port
Measurement active
MessMode_A [sss Aaaa sn.nnnnnn<cr>]
MessMode_B [sss Baaa sn.nnnnnn<cr>]
MessMode_A_minusB [sss Aaaa-Baaa sn.nnnnnn<cr>]
MessMode_A_B [sss Aaaa sn.nnnnnn Baaa sn.nnnnnn<cr>]
sss = 0 .. 255 - Continuous no.
aaa = Sensor Address 1..255 (e.g. 004) / ML/LT NT 1-32
sn.nnnnnn = +9.999999 - Positive Overrange
-9.999999 - Negative Overrange
other value - Angle in rad e.g. +0.226349
Menue aktive or error ( -0.000000 means no data available)
MessMode_A [9xy Aaaa -0.000000<cr>]
MessMode_B [9xy Baaa -0.000000<cr>]
MessMode_A_minusB [9xy Aaaa-Baaa -0.000000<cr>]
MessMode_A_B [9xy Aaaa -0.000000 Baaa -0.000000<cr>]
aaa = Sensor Address 1..255 resp. 0 when address undefined
x = 0 - General error
1 - Port A error
2 - Port B error
y = 0 - Menu active (no error)
>0 - Error codes (According to manual)
Format of transmission
asynchrony, 7Bit, 2 Stopbits, no parity