Page 27 of 31 Pages
4. TECHNICAL DATA
4.1. Power supply
Battery powered 4 Batteries alkaline type, size AA, 2 Ah
or
4 NC rechargeable batteries AA, 0.8 Ah
or
4 Ni-MH- rechargeable batteries, size AA, 1.2 Ah
External power supply
12.0 to 48.0 V, DC, connected to port A or B or OUT
Using external power supply the automatic sleep mode
of the display is inactive.
Remarks:
The rechargeable batteries are automatically charged when an external power supply is connected
The regular batteries can stay in the LEVELMETER 2000 when an external power supply is connected as
they will automatically be shut off.
4.2. Power consumption
LEVELMETER 2000 180 mW, battery powered
500 to 3000 mW, external power supply
Connected sensors / instruments
approx. 100 mW
Transceiver/Converter
1200 mW, only with external power supply
4.3. Battery lifetime
LEVELMETER 2000
with one sensor connected
approx. 40 hours, battery powered
approx. 25 hours, rechargeable batteries used
with two sensors connected
approx. 33 hours, battery powered
approx. 20 hours, rechargeable batteries used (Ni-MH)
4.4. Measuring range
depending on the connected sensors/instruments
4.5. Resolution
depending on the connected sensors/instruments and on the filter type installed
4.6. Various
Dimensions: 100 mm x 210 mm x 45mm (incl. Battery cover)
Weight incl. Batteries: 462 gr.
Temperature range (storage): -20 to +70°C