INSTRUCTIONS FOR pH5 / Cond5 / PC5
POWER ON
Press key once, the meter will switch On and performs a test of 3 colors backlight display and then show
all the segments active for 2 sec.
Then it will display the following:
SETUP MENU for pH5 / Cond5 / PC5
1. With meter switched off, press and hold CAL key and press key once.
2. The meter will switch on with all the segments active, release CAL key, the meter will go into the
Setup Menu (green backlight during setup).
3. Press key to select the parameter to be changed between:
4. Press to enable the value changing of the selected parameter.
5. The value of selected parameter will start flashing.
6. Press to change the value and then press to confirm.
7. The value stops flashing.
8. Press to select other parameters or press ESC to exit the setup menu.
NOTE: Any time press ESC key to exit from SETUP menu.
MEASUREMENT
Rinse the electrode with distilled water or sample prior to start measurement.
Fill the measuring cap with sample, switch on the meter with and press MODE key to select desired
parameter to be measured (green backlight during measurement).
Immerse the Tester in sample and wait for stability, when stability icon ☺ appears on display take the
reading.
During measurement make sure that pH electrode is free from air bubbles, and that there isn’t any air
bubble around or between conductivity sensor.