5
Calibration
In order to ensure the precision of weighing, please preheat the balance completely before the calibration
Procedure
With no load on the plate, keep pressed CAL key for 5 seconds
Scale value blinks for 3 seconds
Put the correspondent external certificated weight on the plate
Display the calibrated valued
Function Operation
To choose the operating mode keep pressed MODE key for 3 seconds:
All functions are displayed, in a cycle:
Press MODE to confirm the selection.
Density measure mode is an option, if you need please contact our company or authorized
Distributor
PCS counting
By this procedure the user measures the total weight of the certain samples which with similar single weight,
then divided by the single sample weight to get the final quantity of the sample.
Procedure
Display asks for sample number
Use Tare to select an alternative number
Put the correspondent numbers of samples on the pan
Press CAL to confirm the sample number
Putt all your samples on the pan
• In Count Mode, to reset the sample number value, press MODE.
• To Exit the Count Mode: Keep pressed MODE key, and then choose the normal weigh mode.