2.1 Before Installation
Selecting the Installation Site (1)
Introduction
In order to make full use of the device functions, please observe the following
criteria when selecting an installation site.
Notes :
Install the product in a location with an ambient temperature of
-25°C - +60°C and an ambient humidity of 5 - 100% RH to prevent
equipment malfunction or output errors.
Avoid installing the product near high-current power lines, motors,
or transformers to prevent damage from electromagnetic induction,
which may cause equipment malfunction or output errors.
DO NOT use this product for grounding a welder, as it may cause
damage to the product.
When welding near this product, be sure to ground the welding
power transformer.
Avoid locations subject to severe vibration or highly corrosive
atmospheres to prevent detector breakage or equipment damage.
Keep the product away from direct sunlight, wind, and rain to pre-
vent output errors.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Yamatake MagneW3000 PLUS and is the answer not in the manual?

Yamatake MagneW3000 PLUS Specifications

General IconGeneral
Repeatability±0.1% of reading
Measurement PrincipleFaraday's law of electromagnetic induction
TypeMagnetic Flowmeter
Fluid Temperature-20°C to +180°C
Lining MaterialPFA, PTFE
Electrode MaterialHastelloy
Accuracy±0.5% of reading
Protection DegreeIP67
Output Signal4 to 20 mA DC, Pulse, HART
Power Supply24 V DC
Measurement Range0 to 10 m/s
Operating Temperature-20°C to +60°C