14
IM700938-01E
Specifications
Product Specifications
Compatible probes* 700939 FET Probe 701930 Current Probe
701931 Current Probe 701932 Current Probe
701933 Current Probe
Number of power connectors
2
Output voltage ±12 V ±0.5 V
Operating temperature and humidity range
0 to 40°C (32 to 104°F), 80% RH or less (no
condensation)
Storage temperature and humidity range
-10 to 50°C (14 to 122°F), 80% RH or less (no
condensation)
Location for use Indoor, altitude up to 2000 m (6562 feet)
Rated output current 600 mA (the total value of both outputs)
Rated supply voltage 100 V AC (120, 220, and 240 V requires specification)
(Voltage fluctuation of 10% from the rated supply
voltage are taken into account.)
Ripple supply voltage 3 mVp-p (at the rated output current)
Rated supply frequency 50/60 Hz
Maximum rated power 60 VA (at the rated output current)
External dimensions Approx. 73(W) × 110(H) × 186(D) mm
Approx. 2.87”(W) × 4.33”(H) × 7.32”(D)
Mass Approx. 1.1 kg
Approx. 38.8 oz.
Standard Accessories Power cord (If the suffix code is -Y, a power cord is
not included.), Instruction manual,
Spare fuse
100 V, 120 V : F1.0 AL/250 V, 20 mm × 5 mm dia.
220 V, 240 V : F0.5 AL/250 V, 20 mm × 5 mm dia.
* For the latest information,
contact your nearest YOKOGAWA
representatives.
Compliant Standards
Safety standards EN61010-1
Pollution Degree 2
EMC standards Emission EN61326-1 Class B
EN55011 Class B, Group 1
EN6100-3-2 Class B
EN6100-3-3
EMC standards of Australia and New Zealand
EN55011 Class B, Group 1
Immunity EN61326-1 Table 1 (Basic immunity requirement)
Environmental standards EU RoHS Directive compliant
Corrective Action to be Taken in the case of an Abnormality
If an message appears on the screen, refer to the following pages.
If maintenance service is required, or if the instrument still does not operate
properly even if the proper corrective action has been taken, contact your
nearest YOKOGAWA representatives, listed on the back cover of this manual.
Symptom Possible Cause Corrective Action
The power cannot be turned ON. The fuse is blown. Replace the fuse.