13-20
IM 04L42B01-01E
TransportandStorageConditions
Item Specifications
Ambient temperature –25 to 60°C
Ambient humidity 5 to 95%RH (no condensation)
Vibration 10 to 60 Hz, 4.9 m/s
2
maximum
Shock 392 m/s
2
maximum (in packaged condition)
SupportedStandards
Item Specifications
CSA CSA22.2 No.61010.1, CSA C22.2 No.61010-2-030, installation category II
*1
, pollution degree 2
*2
,
and measurement category II
*3
UL UL61010-1, UL 61010-2-030 (CSA NRTL/C)
CE
EMC directive EN61326-1 compliance, Class A, Table 2 (For use in industrial locations)
EN61000-3-2 compliance
EN61000-3-3 compliance
EN55011 compliance, Class A, Group 1
Low voltage directive EN61010-1, EN 61010-2-030 compliance, installation category II
*1
, pollution degree 2
*2
, and
measurement category II
*3
EMC regulatory arrangement in Australia and New Zealand
EN55011 compliance, Class A, Group 1
KC marking Electromagnetic wave interference prevention standard, electromagnetic wave protection standard
compliance
WEEE directive Compliant
*1 Installation category (overvoltage category) II: Describes a number which defines a transient overvoltage condition. Iimplies the regulation for
impulse withstand voltage. “II” applies to electrical equipment which is supplied from the fixed installation like a distribution board.
*2 Pollution degree 2: Describes the degree to which a solid, liquid, or gas which deteriorates dielectric strength or surface resistivity is adhering. “2”
applies to normal indoor atmosphere. Normally, only non-conductive pollution occurs.
*3 Measurement category II: Applies to measuring circuits connected to low voltage installation, and electrical instruments supplied with power
from fixed equipment such as electric switchboards.
13.6GeneralSpecifications