EasyManuals Logo

YOKOGAWA Rotamass Supreme User Manual

YOKOGAWA Rotamass Supreme
90 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #89 background imageLoading...
Page #89 background image
Ambient conditions
General Instruction Manual
Specifications
IM 01U10B00-00EN-R, 3rd edition, 2018-07-09
89 / 90
Ranges and specifications
Electromagnetic compatibility (EMC)
IEC/EN 61326-1, Table 2
IEC/EN 61326-2-3
NAMUR NE 21 recommendation
DNVGL-CG-0339, chapter 14
This includes
Surge immunity acc.:
EN 61000-4-5 for lightning
protection
Emission acc.:
IEC/EN 61000-3-2, Class A
IEC/EN 61000-3-3, Class A
NAMUR NE 21 recommendation
DNVGL-CG-0339, chapter 14
Immunity assessment criterion:
The output signal fluctuation is within ±1%
of the output span.
Maximum altitude
2000m (6600ft) above mean sea level
(MSL)
Overvoltage category acc. IEC/EN61010-1 II

Table of Contents

Other manuals for YOKOGAWA Rotamass Supreme

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA Rotamass Supreme and is the answer not in the manual?

YOKOGAWA Rotamass Supreme Specifications

General IconGeneral
BrandYOKOGAWA
ModelRotamass Supreme
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals