EasyManua.ls Logo

YOKOGAWA WT333E - Page 101

YOKOGAWA WT333E
143 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8-9
IM WT310E-01EN
Other Features
1
2
3
4
5
6
7
8
App
Index
1.0 A
0.8 A
0.6 A
5.0 V
4.0 V
3.0 V
5.0 V
Time
Time
Measured current D/A output D/A output
Fixed range mode Manual range mode
(DA zoom)
–5.0 V
Compare (Comparator Mode)
Bycomparingwiththecomparatorlimits,thisinstrumentoutputs+5V,0V,or–5V.Toreplacethe
output with a relay contact output, like the WT210/WT230 comparator function, provide your own
relay and relay driving circuit.
• Belowthelowerlimit: –5Voutput
• Greaterthanorequaltothelowerlimitandlessthantheupperlimit: 0Voutput
• Greaterthanorequaltotheupperlimit: +5Voutput
Maximum and Minimum Values in Manual Range Mode and Comparison Upper and
Lower Limits in Comparator Mode
You can select a value between -9.999E+12 and 9.999E+12.
The default settings are shown below.
ManualRangeMode MaximumValue 100.0
MinimumValue -100.0
Comparator Mode Comparison Upper Limit 100.0
Comparison Lower Limit -100.0
8.4 Configuring D/A Output Items (Option)