EasyManua.ls Logo

YOKOGAWA WT333E - 2.10 Setting the Data Update Interval

YOKOGAWA WT333E
143 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-21
IM WT310E-01EN
Measurement Conditions
1
2
3
4
5
6
7
8
App
Index
2.10 Setting the Data Update Interval
Procedure
Follow the procedure indicated by the thick lines in the following menu.
Select the data update interval.
(Display B)
1.
2.
3.
(Display C)
4.
5.
Confirm
the setting.
Close menu.
6.
5.
An example for the WT333E
The display varies depending
on the installed elements.
9.
Set the synchronization source.
8.
Confirm
the setting.
Close menu.
(Display D)
10.
6.
Set the timeout value.
(Display D)
7.
Explanation
Data Update Interval
The data update interval is the period at which sampling data for determining measurement functions
is acquired.
Youcanselectthedataupdateintervalfromtheoptionsbelow.Thedefaultsettingis0.25s.
0.1s,0.25s,0.5s,1s,2s,5s,10s,20sorAuto
• Ateachdataupdateinterval,thenumericdataisupdated,stored,converted,andoutputasanalog
signals, and transmitted through the communication interface.
• TheUPDATEindicatorblinksinsyncwiththeselectedinterval.
• WhensettoAuto,everytimeaperiod*oftheinputwaveformspecifiedasthesynchronization
source (described later) is detected, the UPDATE indicator blinks, measured data is updated, output
as analog signals, and transmitted through the communication interface.
* 100 ms or more.
• Tocapturerelativelyfastloadfluctuationsinthepowersystem,selectafastdataupdateinterval.To
capture low frequency signals, select a slow data update interval.
• Ifthefluctuationintheinputsignalperiodislarge,selectAuto.