4-4
IM WT310E-01EN
Input Element (WT332E/WT333E only)
The elements that you can select vary depending on the model. Check the model when selecting
elements.
• 1,2,or3:Displaythemeasuredvaluesofelement1,2,or3.
• Σ:Seesection4.1.
Note
• Apowerfactorof1.0001to2.0000isdisplayedas1.0000.Apowerfactorof2.0001orgreaterisdisplayed
asanerror(“Error”).Apowerfactorof–2.0000to–1.0001isdisplayedas–1.0000.Apowerfactorof–
2.0001 or less is displayed as an error (“Error”).
• Evenwhentheinputsignalisthesame,thedisplaymaychangeifyouchangethemeasurementmode
(RMS,VOLTAGEMEAN,orDC).Fordetailsonmeasurementmodes,seesection2.1.
• Wheneitherthevoltageorcurrentfallsto0.5%orless(1%orlessifthecrestfactorissetto6or6A)of
the measurement range, the power factor will be displayed as an error (“Error”).
4.2 Displaying Apparent Powers, Reactive Powers, and Power Factors