YSI Incorporated Model 5000/5100 Service Manual 2
SECTION 2 SPECIFICATIONS
OXYGEN MEASUREMENT
Ranges: mg/L: 0.0 to 60.0 mg/L
% air saturation: 0.0 to 600.0% air saturation
mbar: 0.0 to 1500
Accuracy: mg/L: ±0.1% plus 1 Least Significant Digit (LSD)
% air saturation: ±0.1% plus 1 LSD
mbar: ±1% plus 1 LSD
Resolution: mg/L: 0.1%, or 0.01 mg/L , whichever is greater
% air saturation: 0.1% air saturation
mbar: 1 mbar
TEMPERATURE MEASUREMENT
Range: -5.0 to +50.0°C
Accuracy: ±0.1°C
Resolution: 0.01°C
BAROMETRIC PRESSURE MEASUREMENT
Range: 450 to 825 mmHg (600 to 1100 mbar)
Accuracy: ±1% plus 1 LSD within ±10°C ambient temperature from calibration point
Resolution: 1 mmHg
TEMPERATURE COMPENSATION
The mg/L mode is automatically temperature-compensated to an accuracy of ±1% of DO
readings between 0 and 5°C, and to an accuracy of ±0.6% of readings between 5 and 45°C.
The % air saturation mode is automatically temperature-compensated to an accuracy of ±0.5% of
calibration values between 0 and 5°C, and to an accuracy of ±0.3% of values between 5 and
45°C.
SALINITY COMPENSATION
Range: 0.0 to 40.0 ppt
Accuracy: ±.02 mg/L
OPERATING ENVIRONMENT
0 to 45°C, 10 to 90% relative humidity, non-condensing
WATER RESISTANCE
The Model 5000 and 5100 are designed exclusively for indoor use and are NOT waterproof.
POWER
The Model 5000 & 5100 are powered by an AC adapter or 4 C-size alkaline batteries. A new set
of alkaline batteries will power the selected instrument for approximately 30 hours (not including
stirring).