Description of Menu Parameters
(1) Input Signal Type – tyPE: There are corresponding signs for thermal
resistances and thermocouples.
(2) Start-up Customization – ConFig: If the current mathematic measurement
mode, display unit, resolution and reference end compensation mode are expected
to be kept for each start-up in the future, then the menu parameter “ConFig”
should be set to 1, so that the current measurement and display forms can be saved
after exiting the menu (the ConFig return to zero automatically at the same time).
(3) Automatic Shut-off Time – AutoFF: Its value represents the automatic
shut-off time (minutes) in case of no key-press, and the function of automatic
shut-off will be canceled in case of AutoFF=0.
(4)Automatic Backlight Shut-off Time - Auto.b.L: Its absolute value is the
automatic backlight shut-off time (second), the automatic backlight shut-off
function will be canceled in case of Auto.b.L=0, and the negative sign only means
that the buzzer sound for key-press and alarm is not allowed.
(5) Instrument Calibration Code - CAL.Cod: User calibration is allowed in
case of CAL.Cod=808; it will restore the factory calibration datum and settings in
case of CAL.Cod=5500.
(8) Sensor Correction Value (Offset) - OFF.SEt: The measured value is
translated in the form of electric quantity value, i.e. x
i
=x
i
+ OFF.SEt. Only the
electric quantity value can be used for setting.