Cleaning / Disinfection
8
Page 17
General notes
Before starting maintenance and cleaning measures, the device must always
be switched off at the main switch (18) and the power cable must be
disconnected (17).
During cleaning and disinfection, make sure that the labels of the device (such
as warnings, control unit signs, identification plate) are not damaged.
During cleaning and disinfection, make sure that no liquids penetrate the
device. Do not use sprays.
If liquid gets into the device during cleaning or disinfection, take the device out
of use, ensure that it cannot be used again, and contact your distributor.
Use the device only in a proper environment.
Housing /
accessories
Cleaning: If there is visible soiling, the housing, cable, display and accessories
can be cleaned with commercially available alcohol-free plastic cleaners. Wipe
the surface until the dirt has been removed using a soft cloth which has been
soaked according to the information from the manufacturer of the cleaning
agent but not dripping wet.
Disinfection: We recommend performing disinfection at least once weekly and
if there are any signs of contamination. Contact your hygiene specialist if
necessary. Always perform cleaning prior to disinfection. Housing, cables and
accessories can be disinfected using disinfectant wipes. Use a commercially
available alcohol-free disinfectant for metal and plastic, with bactericidal,
virucidal, and fungicidal properties. Observe the manufacturer’s instructions for
use. Wipe all surfaces with a soft cloth which has been soaked according to the
disinfectant manufacturer’s instructions but which is not dripping wet, or use
pre-soaked disinfectant cloths (wipes).
Also observe the instructions for drying or post-cleaning, where applicable.
Defrost container
Cleaning: The defrost container should be cleaned whenever it has been
emptied.
To do this, you can open the lid of the defrost container. Follow the procedure
described under “Housing / Accessories”.
Air filter
To filter coarse dust particles from room air, there is an air filter located at the
back of Cryo 7.
If the icon to change the air filter appears in the status bar, the air filter along
with the frame must be changed.
The Cryo 7 air filter cannot and should not be cleaned.
Notes
Use the device only in a properly hygienic environment.