11
7. Result Overview
1500µg/ml
1500µg/ml
10 Mins
The reader will display the result in μg/ml on the
screen after 10 minutes (in Standard Mode) or
after the test cartridge has been inserted after
exactly 10 minutes have passed (in Quick Mode).
The detection range is between 0 - 3000 μg/ml,
where zero means SAA is not detected.
1
0
20 300
HIGH SAA
MODERATE SAA
LOW SAA
mins minsmins
Development of test results over time
Test cartridges must be read by the EQ-1
reader at exactly 10 minutes because the
cartridge Control and Test lines continue to
develop over time.
This is especially important in Quick Mode as
the test is timed by you, independent of the
reader.
Inserting a cartridge beyond the 10-minute
mark will display a falsely elevated SAA
value.
Conversely, inserting a cartridge before
10 minutes have passed, will read a lower
SAA value than is present.
Important: Re-inserting the test cartridge
beyond the 10-minute mark, to recheck the
result is not recommended and will give an
erroneous, falsely elevated result.