4
9. Each day that a processed 1295BI is incubated,
activate and incubate at least one non‑processed
1295BI to use as a positive control. Follow the
activation instructions provided in Step 8above.
Write a "C" (for "control") and the date on the BI
label. The positive control should be from the same
lot code as the processed biological indicator. The
positive control BI helpsconfirm:
• correct incubation temperatures aremet,
• viability of spores has not been altered due
to improper storage temperature, humidity or
proximity tochemicals,
• capability of media to promote rapid growth,and
• proper functioning of the Auto‑reader.
10. Incubation andReading:
Incubate the positive control and processed
1295BIs at 60± 2°C in a 490H Auto‑reader or
490Auto‑reader having software version 4.0.0or
greater. See the Auto‑reader Operator’s Manual for
the proper use of this equipment. The Auto‑reader
will indicate a positive result as soon as it is obtained.
The final negative 1295BI reading is madeat:
‑ 24minutes in 490H or 490Auto‑readers having
software version 4.0.0orgreater
‑ 4hours in 490H Auto‑readers having a software
version less than4.0.0.
After the results are displayed and recorded, the
1295BIs may bediscarded.
FluorescentResult
The positive control (unprocessed) 1295BI must
provide a positive fluorescent result (+ symbol on
the 490H Auto‑reader or 490Auto‑reader having
software version 4.0.0or greater LCD display).
Processed 1295BI results are not valid until the
positive control reads fluorescent positive. If the
positive control reads negative (‑ symbol on the LCD
display), check the Auto‑reader Operator’s Manual
Troubleshooting Guide. Retest the 490H Auto‑reader
or 490Auto‑reader having software version 4.0.0or
greater with a new positivecontrol.
With processed 1295BIs, a final negative reading
(‑ symbol on the LCD display) indicates an acceptable
sterilization process. A positive (+ symbol on the LCD
display) result for a processed 1295BI indicates a
sterilization process failure. Act immediately on any
positive results for processed BIs. Determine the
cause of the positive BI following facility policies and
procedures. 3M recommends retesting the sterilizer
according to your facility policy before processing
additionalloads.
Optional Visual pH Color ChangeResult
The 1295BI is normally discarded after the fluorescent
result has been recorded. If, however, special studies
are desired, 1295BIs may be further incubated for
a visual pH color change result. In the case of the
positive control BI, a yellow color change of the growth
media will appear within 24hours. Any observation of
a yellow color within the vial indicates a positiveresult.
In the case of a processed 1295BI, a media color
change from purple to yellow indicates a sterilization
process failure. A negative pH color change result,
i.e., media remains purple, can be assessed at 7days.
To avoid media dry‑out, it is recommended that the
1295BI be transferred to a humidified incubator
operating at 60ºC after the fluorescent result has
beenrecorded.