Bair Hugger Model 775 Temperature Management Unit Operator’s Manual
4
Arizant Healthcare Inc.
Overview and Operation
Unit Power-On-Reset
The Model 775 temperature management unit automatically performs the following
power-on-reset sequence after you connect the unit to a properly grounded power source:
Performs all self-test functions.
Illuminates indicator lights and all pixels in the alphanumeric display momentarily.
Displays the text “MD 700’s” and the software revision level in the alphanumeric
display.
Sounds the alarm (three low-level clicks).
Enters the Standby mode with the high fan setting selected by default.
Note: e low fan setting indicator light will not illuminate during power up.
If the temperature management unit looses power for less than 1 second, the units software
will return the unit to the operating mode you selected prior to the power loss. However, if
the temperature management unit loses power for longer than 1 second, the units software
will reset when you restore power. The unit will then enter the Standby mode.
Figure A. Model 775 temperature
management unit front view
Figure B. Model 775 temperature
management unit – rear view
Carrying handle
Alphanumeric
display
Control Panel
Hose connection
Power cord
connection
Hose
IV pole clamp