7
Bair Hugger Model 775 Temperature Management Unit Operator’s Manual
English 771022D
Mounting the Temperature Management Unit to an IV Pole
caution
To prevent tipping, clamp the Model 775 temperature management unit to an IV pole
at a height that provides stability. We recommend clamping the unit no higher than 44"
(112 cm) from the floor on an IV pole with a minimum 28" (71 cm) diameter wheelbase.
Failure to do so may result in IV pole tipping, catheter site trauma and patient injury.
method
To mount the Model 775 temperature management unit to an IV pole, simply turn the
clamp handle clockwise to tighten the clamp on the IV pole and counterclockwise to
release it. Ensure the IV pole bumper rests against the pole. See Figure D.
28 in. (71 cm) minimum
44 in.
(112 cm)
maximum
Figure D. Model 775 on an IV Pole
IV Pole
Bumper
Clamp
Handle