Bair Hugger Model 775 Temperature Management Unit Operator’s Manual
12
Arizant Healthcare Inc.
Specifications
Physical Characteristics
 13" high x 13" deep x 14" wide
33 cm high x 33 cm deep x 36 cm wide
 16 lb (7.3kg)
  53 dBA (High fan setting)
48 dBA (Low fan setting)
 Flexible, compatible with the Bair Hugger brand 241
blood/fluid warming system.
  High efficiency 0.2 µm air filter
   Every 12 months or 500 hours of use.
 Can be clamped to an IV pole, placed on a hard surface,
or mounted to the rolling stand accessory.
Temperature Characteristics
   15°C-25°C
 Electronically controlled.
  High fan setting: 1600 BTU/hr (average), 470 W (average)
Low fan setting: 1330 BTU/hr (average), 390 W (average)
  Average temperatures at the end of the hose:
HIGH: 43° ± 1.5°C 109.4° ± 2.7°F
MED: 38° ± 1.5°C 100.4° ± 2.7°F
LOW: 32° ± 1.5°C 89.6° ± 2.7°F
Safety System
 Independent electronic circuit; thermal cutoff shuts the
heater OFF to ensure hose end air remains below 56°C
(5C ± C typical); back-up over-temperature detection
at hose inlet.
  Over-temperature (≤56°C, 53°C ± 3°C typical): red Over-
temp indicator light flashes, alarm sounds, heater and
blower shut down, operating indicator lights turn OFF,
control panel becomes unresponsive.
Fault: amber Fault indicator light flashes, alarm sounds.
 Dual input fused lines.