Bair Hugger Model 775 Temperature Management Unit Operator’s Manual
6
Arizant Healthcare Inc.
standby
Upon power up, the warming unit defaults to Standby and the high fan setting. When the
unit is in 32˚C (Low), 38˚C (Med), or 43˚C (High) or Ambient mode, press the Standby
button to place the unit in Standby mode.
When you select the Standby mode, the following events occur:
• Standby indicator light illuminates.
• Blower and heater are turned OFF.
• Alphanumeric display deactivates.
• Alarm and fault detection functions remain active.
• Operating timer pauses.
• Airow mode resets/defaults to the high fan setting.
temp in range indicator light
The Temp in Range indicator light illuminates when the temperature at the blanket
or gown end of the hose is ±1.5°C of the selected setting; this indicator light does not
illuminate in the Ambient mode.
fault indicator light
When a system fault occurs, the amber Fault indicator light flashes and an alarm
sounds.
Please refer to What to Do in Case of a Fault Condition on page 9 for additional
information.
over-temp indicator light
If the unit senses an over-temperature condition, the red Over-temp indicator light
flashes and an alarm sounds.
Please refer to What to Do in Case of an Over-Temperature Condition on page 9 for
additional information.