BM series 57
Title block and end block
When mass values are recorded as GLP data, a "Title block" is inserted at the beginning and an
"End block" is inserted at the end of a group of mass values in the GLP report.
Notes
To output the report to an AD-8121B , use MODE 3 of the AD-8121B. If MODE1 is used, select
temporary dump print mode by pressing the STAT. key of the AD-8121B.
If the data memory function is used (except data 0), the "Title block" and "End block" can not be
output.
Key operation
1 With the weighing data displayed, press and hold the PRINT key, then release the key.
5tart is displayed. The "Title block" is output.
2 The weighing data is output according to the parameter setting of the data output mode (prt) of
the function table.
3 Press and hold the PRINT key until recend is displayed, then release the key.
The "End block" is output.
Setting of "info 1" Setting of "info 2"
AD-8121 printer format General format
Space, ASCII 20h
<TERM>
Terminator, CR , LF or CR
CR Carriage return, ASCII 0Dh
LF Line feed, ASCII 0Ah
-----------A-&-D
MODEL-----BM-300
S/N-----01234567
ID------LAB-0123
DATE--2011/02/20
START
TIME----12:34:56
WT--+123.4567--g
WT--+123.4612--g
WT--+123.4623--g
WT--+123.4634--g
WT--+123.4511--g
WT--+123.4534--g
WT--+123.4712--g
WT--+123.4645--g
END
TIME----12:40
15
SIGNATURE
---------------
Title block
Manufacturer
Model
Serial number
ID number
Date
Time
Weighing data
Time
Signature
End block