EasyManua.ls Logo

A&D GP-40K - 11-2 Memory for Weighing Data

A&D GP-40K
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
57
Title block and end block
When weight values are recorded as GLP data, a “Title block” is inserted at the beginning and an
“End block” is inserted at the end of a group of weight values in the GLP report.
Note
To output the report to an AD-8121B , use MODE 3 of the AD-8121B.
Caution
If the data memory function is used, the “Title block” and “End block” can not be output.
Operation
1 With the weighing data displayed, press and hold the PRINT key, then release the key.
S5tart is displayed. The “Title block” is output.
2 The weighing data is output according to the parameter setting of the data output mode.
3 Press and hold the PRINT key until recend is displayed, then release the key. The “End
block” is output.
When the setting is “info 1”: When the setting is “info 2”:

Table of Contents