EasyManua.ls Logo

A&D HR-250A - Page 41

A&D HR-250A
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
39
Class Item and Parameter Description
0
Key mode
Accepts the PRINT key only when
the dis
p
la
y
is stable.
1
Auto print mode A
(Reference = zero)
2
Auto print mode B
(Reference = last stable value)
Outputs data when the display is
stable and conditions of
ap-p
,
ap-b
and the reference value are met.
3
Stream mode
Outputs data at the specified
display refresh rate.
e.g.: Used when connected to the
AD-8920A remote display.
4
Key mode B (Immediately)
Accepts the PRINT key regardless
of the display condition.
5
Key mode C (When stable)
Accepts the PRINT key immediately
when the display is stable, or waits
for the display to be stable when not.
prt
Data output mode
6
Interval output mode
Uses interval output mode.
0
Plus onl
y
Dis
p
la
y
ed value>Reference
1
Minus onl
y
Dis
p
la
y
ed value<Reference
ap-p
Auto print polarity
2
Bo
t
h
Re
ardless of dis
la
ed value
0
10 di
g
its
1
100 di
g
its
ap-b
Auto print difference
2
1000 di
g
its
Difference between reference value
and displayed value
0
Ever
y
dis
p
la
y
refresh
1
2 seconds
2
5 seconds
3
10 seconds
4
30 seconds
5
1 minute
6
2minute
7
5minute
int
Interval time
8
10 minute
Interval time for the interval output
mode
(With
prt 3
)
0
No out
p
ut
1
Time onl
y
3
Date onl
y
5-td
*
Time/Date output
4
Time and date
0
No out
p
ut
5-id
ID number output
1
Out
p
ut
Selects whether or not the ID
number is output.
0
No
p
ause
pU5e
Data output pause
1
Pause
(
1.6 seconds
)
Selects the data output interval.
0
Not used
at-f
Auto feed
1
Used
Selects whether or not automatic
feed is performed.
0
No out
p
u
t
1
AD-8121 forma
t
info
GLP output
2
General data forma
t
Selects the output format for the
GLP/GMP compliant report.
0
Not used
dout
Data output
ar-d
Zero after output
1
Used
Adjusts zero automatically after data
is output
0
600 bps
1
1200 bps
2
2400 bps
3
4800 bps
4
9600 bps
5if
Serial
interface
bp5
Baud rate
5
19200 bps
Factory setting Digit is a unit of minimum weighing value
*
:
Only for the HR-AZ series