UC-321PL UC321-BT UC-321
CALIBRATION INSTRUCTIONS (Ver. 5, 6, 7 Software)
Note: Below, position Weight A kg refers to the switch position furthest away
from the test/data socket. Weight A lb refers to the switch position next to
Weight A kg.
1. NO calibration Jack is required for this version 5,6 and 7 UC321.
2. Set the mode switch on the bottom of the UC321 to the position furthest
away from the test/data socket.
3. Press the Blue Measure Switch once then slide the mode switch towards
the test/data socket, (all the way) and then back again. Within two seconds.
4. If the display shows “--- - - ”, press Measure Switch three times.
5. After the display turns off then back on the display will show “- - - - - again.
6. Press Measure Switch momentarily. The display will show “99999
9999999999
99999” and start
to count down. Press Measure Switch. The display will show0
00
0
7. To check the calibration type set the mode switch to the position furthest
away from the test/data socket and press the measure switch.
By default, scale is set to l
ll
l-
--
-1
11
1, which uses 3 point calibration.
To change set mode switch to the next position towards the test/data socket
then press Measure Switch momentarily. Use Measure Switch to change
setting.
l
ll
l-
--
-0
00
0:2 point, (0kg,150kg).
l
ll
l-
--
-1
11
1:3 point, (0kg,80kg,150kg). * DEFAULT.
l
ll
l-
--
-2
22
2:2 point, (0lb,350lb).
l
ll
l-
--
-3
33
3:3 point, (0lb,176.4lb(80kg),350lb).
Slide Mode switch to position furthest away from the test/data socket press
Measure Switch repeatedly to show
5
55
5
8. Press Measure Switch momentarily, the display will show the temperature
at calibration; i.e. (20
2020
20). This must be set before calibration.
9. To change set mode switch to Weight A lb position then press Measure
Switch, press Measure Switch momentarily, the temperature value will
increase one digit at a time. When the correct temperature is displayed slide
Mode switch to Weight A kg press Measure Switch display “6
66
6
10. Press Measure Switch momentarily, the display will show the gravity
setting applied at time of calibration, i.e. (9798
97989798
9798).
11. To change set mode switch to Weight A lb position then press Measure
Switch,
12. Press Measure Switch momentarily to increment the value one digit at a
time. Once the correct value is displayed, slide Mode switch to Weight A kg
press Measure Switch press Measure Switch to display “7
77
7
13. Press Measure Switch momentarily, the display will show the gravity
setting for the location of use of the scale, i.e. (9798
97989798
9798). If the scale is to be
used at the same location, where it is calibrated, this value must be set
exactly the same as above!
14. Use the same procedure as above to set the gravity value.
……..cont. page 2

Other manuals for A&D UC-321PL

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the A&D UC-321PL and is the answer not in the manual?

A&D UC-321PL Specifications

General IconGeneral
ModelUC-321PL
CategoryScales
DisplayLCD
Auto-offYes
Dimensions320 x 320 x 57 mm
Unitslb / kg / st
Operating Temperature10°C to 40°C
Weight2 kg
Power Source4 x AA batteries

Related product manuals