UC-321 BT CALIBRATION INSTRUCTIONS .cont. (Ver. 5, 6, 7 Software)
15. Press Measure Switch to display “10
1010
10”.
16. Press Measure Switch momentarily, the display will show C
CC
Cl
ll
l 0
0 0
0, remove all
weight from the platform and position the timber calibration jig. Wait until
the stable indicator appears.
17. Press Measure Switch for about half a second, the display will show “11
1111
11”
18. If using 3 point calibration, l
ll
l-
--
-1
11
1 or l
ll
l-
--
-3
33
3, apply 80 kg on top of the calibration
jig.
If using 2 point calibration, l
ll
l-
--
-0
00
0 or l
ll
l-
--
-2
22
2, apply 150 kg on top of the calibration
jig.
Press Measure Switch momentarily, the display will show C
CC
Cl
ll
l1
11
1. Wait until
the stable indicator appears.
19. Press Measure Switch for about half a second, the display will show “12
1212
12”.
20. If using 2 point calibration, l
ll
l-
--
-0
00
0 or l
ll
l-
--
-2
22
2, Weight A lb position then back to
Weight A kg position, proceed to step 22.
If using 3 point calibration, l
ll
l-
--
-1
11
1 or l
ll
l-
--
-3
33
3, apply 150 kg on top of the calibration
jig, and then Press Measure Switch momentarily, the display will show
Cl
ClCl
Cl
2
22
2. Wait until the stable indicator appears.
21. Press Measure Switch for about half a second, the display will show “13
1313
13”.
22. Press Measure Switch repeatedly to display “1
11
15
55
5”
23. Remove weight from calibration jig then press Measure Switch
momentarily, the display will show the weight in kg continuously in this
mode. Apply weights to test performance.
24. Press and hold Measure Switch until the display turns off.