166 AO2000 CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
Softkeys
The following softkeys are displayed in the menu depending on which op-
tions exist for display or configuration of a particular component:
DELETE COMP.
The component can be deleted (see page 175).
CHANGE COMP.
The configuration of the component can be changed (see page 172).
NEW COMP.
A new component can be added (see page 168).
UNIT SELECTION
The physical unit used for the measured value display can be changed.
DETAIL
The detailed display, i.e. a separate window with a detailed list of the com-
ponent data, is opened (see below).
OVERVIEW
The detailed display is closed.
Detailed display
The following component data are listed in the detailed display:
Display
Name of the component in the measured value display, e.g. structure formula
Name
Full name of the component
Molar mass g/mol
Molar mass in g/mol
Response Factor
Detector reply related to propane for the respective component
Quantity of C-atoms
Number of C-atoms in the molecule of the component
Strip Factor
Ratio of the concentration of a component in the water to the concentration
of the same component in the gas flow after the strip event
Unit
The physical units used for readout values