AQUAMASTER4 | ELECTROMAGNETIC FLOWMETER INSERTION SENSOR | OI/FEW400/AP-EN REV.B
…5 Installation
Potting the remote insertion sensor terminal housing
WARNING
Power up and check all connections before
potting.
Pot the remote insertion sensor terminal box
connections immediately on completion (after
power-up and check) to prevent the ingress of
moisture.
Potting materials are toxic. Read the
manufacturers' instructions carefully before
preparing the potting material and use suitable
safety precautions.
Do not overfill or allow the potting material to
come into contact with 'O' rings or grooves.
Do not let potting material enter conduit (if used).
Figure 18 Potting the remote insertion sensor terminal housing
Pipeline preparation
CAUTION
Prepare the main or pipeline into which this
insertion sensor is to be installed prior to
installation.
Installation must be performed only by suitable
experienced/qualified personnel in accordance
with relevant safety requirements.
Safety mechanism
WARNING
The insertion sensor is provided with a safety
mechanism that must be attached to its securing
collar
A as shown in Figure 19 (Secured position).
This prevents rapid outward movement by the
insertion sensor if nut
B is released.
To ensure maximum safety, the positioning collar
C MUST be tightened in place using a 4 mm
hexagon key.
Unsecured
position
Secured
position
See
Warning
Figure 19 Safety mechanism