EasyManua.ls Logo

ABB C13

ABB C13
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2CMC486004M0201 11 C13
Revision: A User Manual
Installation
3.2 Environmental Considerations
Ingress protection
To comply with the protection requirements the product must be mounted in pro-
tection class IP 51 enclosures, or better, according to IEC60259.
Mechanical environment
In accordance with the Measuring Directive (2014/32/EU), the product complies
with M1, which means that it can be operated in “...locations with vibration and
shocks of low significance, e.g. for instruments fastened to light suporting struc-
tures subject to negligible vibrations and shocks transmitted from local blasting
or pile-driving activities, slamming doors, etc.”
Electromagnetic environment
In accordance with the Measuring Directive (2014/32/EU), the product complies
with E2, which means that it can be operated “...in locations with electro magnetic
disturbances corresponding to those likely to be found in other industrial build-
ings.”
Climatic environment
In order to work properly the product should not be operated outside the specified
temperature range of -25°C – +70°C.
In order to work properly the product should not exposed to humidity exceeding
the specified 75% yearly average, 95% on 30 days/year.

Other manuals for ABB C13

Related product manuals