106 CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E
… 9 Operation
… Parameter descriptions
Menu: Display
Menu / parameter Description
Display
Language Selection of menu language. Available languages:
English, Deutsch, Français, Español, Português, Italiano, Chinese
Contrast Contrast setting for the LCD display.
...Operator Pages Selection of submenu ‘...Operator Pages’ using .
Up to four user-specific operator pages (layouts) can be configured for the process display. If multiple operator pages have
been configured, these can be scrolled through manually on the information level. In the factory setting only Operator Page
1 is enabled.
Autoscroll If Autoscroll is enabled, the ‘Autoscroll’ function can also be activated on the information level of the operator menu.
In this function, operator pages are automatically displayed in succession on the process screen, changing every 10
seconds. Manual scrolling through pre-configured operator pages as described above is no longer necessary. When auto
scroll mode is enabled, the  symbol appears in the lower left-hand corner of the display.
Default setting: Disabled.
Mass Flow Format Selection of number of decimal places (maximum 12) used to display the corresponding process variables.
Mass Format
Volume Flow Format
Volume Format
Temperature Format
Density Format
Concentration Format
Date / Time Format Set the display format for the date and time.
Display Test Start the test of the LCD display with ‘ ’. The display test lasts approx. 10 seconds. Various patterns are shown on the
LCD display to check the display.
Display / ...Operator Pages
...Operator Page 1 Selection of submenu ‘...Operator Page 1’ using .
...Operator Page 2 Selection of submenu ‘...Operator Page 2’ using .
...Operator Page 3 Selection of submenu ‘...Operator Page 3’ using .
...Operator Page 4 Selection of submenu ‘...Operator Page 4’ using .
Display / ...Operator Pages / ...Operator Page 1 (n)
Display Mode Configure each operator page. The following variants can be selected:
Off, Graph View, 1x4, 1x6A, 1x6A bar, 1x9, 1x9 bar, 2x9, 2x9 bar, 3x9.
Selecting ‘Off’ deactivates the corresponding operator page.
1st Line Selection of process variable displayed in the respective row.
See table Available process variables on page 80.
2nd Line
3rd Line
4th Line
Bargraph Selection of process variable displayed as a bar graph.
Mass Flow [%]: Mass flow in %
Volume Flow [%]: Volume flow in %