CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E 137
Error code / error text Description NAMUR classification
M038.09
Sensor memory defective.
SensorMemory in frontend board defective.
Check if the SensorMemory is defective.
Contact ABB Service.
Maintenance required
M037.014
NV chips defect on Motherboard.
SensorMemory in motherboard faulty.
Check if the SensorMemory is defective.
Contact ABB Service.
Maintenance required
M032.022
Curr.Out 31/32 not calibrated.
Current output 31 / 32, Uco not calibrated.
Contact ABB Service.
Maintenance required
M031.023
Curr.Out V1/V2 not calibrated.
Current output (plug-in card) V1 / V2 or V3 / V4 not calibrated.
Check plug-in card and replace if necessary.
Contact ABB Service.
Maintenance required
M030.024
Curr.Out V3/V4 not calibrated.
Maintenance required
M028.007
Display value is <1600h at Qmax.
The current counter reading has exceeded the display resolution.
Check the unit setting for the mass counters / volume totalizers and adjust if necessary.
Maintenance required
M026.004
Maintenance interval is reached
Maintenance interval reached.
• Perform maintenance work.
Start new maintenance interval in the ‘Diagnostics / ...Diagnosis Control’ menu.
Maintenance required
M024.008
Device not calibrated.
Contact ABB Service. Maintenance required
M020.027
Communicat. card not responding
Fieldbus plug-in is not reacting.
Plug-in card is defective.
Contact ABB Service.
Maintenance required